Deprecated: Implicit conversion from float 211.6 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 211.6 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 211.6 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 211.6 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 211.6 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 245.2 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 245.2 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 245.2 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534
Deprecated: Implicit conversion from float 245.2 to int loses precision in C:\Inetpub\vhosts\kidney.de\httpdocs\pget.php on line 534 Am+J+Clin+Pathol 2021 ; 155 (6): 773-775 Nephropedia Template TP
gab.com Text
Twit Text FOAVip
Twit Text #
English Wikipedia
Unexpected False-Positive Rates in Pediatric SARS-CoV-2 Serology Using the EUROIMMUN Anti-SARS-CoV-2 ELISA IgG Assay #MMPMID33899091
Geisler D; Freeman MC; Rapsinski GJ; Wheeler SE
Am J Clin Pathol 2021[May]; 155 (6): 773-775 PMID33899091show ga
OBJECTIVES: Serologic assay performance studies for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-?2) in pediatric populations are lacking, and few seroprevalence studies have routinely incorporated orthogonal testing to improve accuracy. METHODS: Remnant serum samples for routine bloodwork from 2,338 pediatric patients at UPMC Children's Hospital of Pittsburgh were assessed using the EUROIMMUN Anti-SARS-CoV-2 ELISA IgG (EuroIGG) assay. Reactive cases with sufficient volume were also tested using 3 additional commercial assays. RESULTS: Eighty-five specimens were reactive according to the EuroIGG, yielding 3.64% (95% confidence interval [CI], 2.91%-4.48%) seropositivity, of which 73 specimens had sufficient remaining volume for confirmation by orthogonal testing. Overall, 19.18% (95% CI, 10.18%-28.18%) of samples were positive on a second and/or third orthogonal assay. This 80.82% false positivity rate is disproportionate to the expected false positivity rate of 50% given our pediatric population prevalence and assay performance. CONCLUSIONS: In pediatric populations, false-positive SARS-CoV-2 serology may be more common than assay and prevalence parameters would predict, and further studies are needed to establish the performance of SARS-CoV-2 serology in children.