  lüll Pacemaker malfunction Hayes DL; Vlietstra REAnn Intern Med  1993[Oct]; 119 (8): 828-35The field of cardiac pacing has expanded rapidly in recent years. Engineering  improvements and microprocessor technology have resulted in a vast increase in  pacemaker technology. The "high-tech" edge of cardiac pacing often discourages  all but the pacemaker specialist from approaching a malfunctioning pacemaker.  Electrocardiographic signs of pacemaker malfunction can be grouped into four  categories: failure to output, failure to capture, undersensing, and  inappropriate pacemaker rate. For each of these categories, there may be true  malfunctions and pseudomalfunctions. In addition, environmental sources of  electromagnetic interference, both within and outside the hospital environment,  can result in pacemaker malfunction. Approaching pacemaker malfunction with these  categories in mind should help minimize confusion.|*Pacemaker, Artificial[MESH]|Electrocardiography[MESH]|Environmental Exposure[MESH]|Equipment Design[MESH]|Equipment Failure[MESH]|Humans[MESH] |